Fig. 4-31
Transport motor drive
The bridge drives M1, the encoder reports its speed, and R1 turns current into a voltage you can read. Two of those three readings together tell the whole story.
Test point
Expected
24 V⎓ ±1
Red to TP6, black to TP5 on the GND bus. This is the rail the bridge switches.
Tap a test point
Reading the print
- +24 V arrives through F1 and lands on the bridge at TP6.
- The bridge chops that rail into PWM for M1; the average voltage at TP9 sets the speed.
- R1 turns motor current into a voltage: 50 mV per amp, so 200 mV is 4 A.
- The controller watches TP7 to hold commanded speed. No tach, no speed control.
- A dragging belt shows high current and no tach; a shorted bridge shows full voltage at rest.
When it does not match
Belt crawls under a light load
Check TP8 high, TP9 low, TP7 steady
Mechanical drag or a failing winding. Compare current against a known-good belt before you swap the motor.
Motor runs flat out as soon as power is applied
Check TP9 with the controller disconnected
A shorted bridge device passes the rail straight through. Replace the bridge, then find what killed it.
Motor hums and does not move
Check TP9 in both directions
One half of the bridge is not switching. The winding is fine; the drive is not.
Speed wanders, tach reading jumps
Check TP7 while wiggling the motor connector
Intermittent tach wiring reads as speed error. Reseat the connector before replacing the motor.