All prints

Fig. 4-20

Power distribution and fusing

Start here whenever the panel is dead or a fuse keeps opening. Every other sheet depends on these rails being right.

J1 AC inletFL1 line filterT1 transformerPS1 regulatorF1 / F2 fuses
J1AC INFL1FILTERT1120:24PS1REG+24 V+12 V+5 VF1 5AF2 2AM1TRANSPORTK1SOLENOIDA1CONTROLGND BUSTP1TP2TP3TP4TP5DW-100 · SERVICE PRINTSFig. 4-20 · Power distribution and fusing

Test point

TP1AC line in

Expected

120 V~ ±6

dial V~red → VΩblack → COM

Across J1 L and N with the machine plugged in and the switch on.

Live mains. One hand in the machine, fingers behind the barriers.

Reading the print

  • J1 feeds FL1, and the filtered mains land on T1's primary.
  • T1 drops 120 V~ to 24 V~; PS1 rectifies and regulates that into +24, +12 and +5 V⎓.
  • F1 (5 A) sits in the +24 V rail ahead of the transport motor.
  • F2 (2 A) protects the +5 V rail and everything on the control board.
  • The GND bus is the reference for every reading on the other two sheets.

When it does not match

Panel dead, no display

Check TP1, then TP4

No mains at TP1 means the cord or the inlet. Mains present but no 5 V points at F2 or PS1.

F1 opens as soon as it is replaced

Check TP2 with the motor unplugged

Find the short before you fit another fuse. Never fit a larger one to make it stay.

Display resets when the transport starts

Check TP4 while jogging the belt

A 5 V rail that dips under 4.75 V under load is a tired filter cap on PS1.

120 V at the inlet, nothing downstream

Check TP1 at both ends of FL1

An open filter or a burnt transformer primary stops everything after it.