Fig. 4-20
Power distribution and fusing
Start here whenever the panel is dead or a fuse keeps opening. Every other sheet depends on these rails being right.
Test point
Expected
120 V~ ±6
Across J1 L and N with the machine plugged in and the switch on.
Live mains. One hand in the machine, fingers behind the barriers.
Tap a test point
Reading the print
- J1 feeds FL1, and the filtered mains land on T1's primary.
- T1 drops 120 V~ to 24 V~; PS1 rectifies and regulates that into +24, +12 and +5 V⎓.
- F1 (5 A) sits in the +24 V rail ahead of the transport motor.
- F2 (2 A) protects the +5 V rail and everything on the control board.
- The GND bus is the reference for every reading on the other two sheets.
When it does not match
Panel dead, no display
Check TP1, then TP4
No mains at TP1 means the cord or the inlet. Mains present but no 5 V points at F2 or PS1.
F1 opens as soon as it is replaced
Check TP2 with the motor unplugged
Find the short before you fit another fuse. Never fit a larger one to make it stay.
Display resets when the transport starts
Check TP4 while jogging the belt
A 5 V rail that dips under 4.75 V under load is a tired filter cap on PS1.
120 V at the inlet, nothing downstream
Check TP1 at both ends of FL1
An open filter or a burnt transformer primary stops everything after it.